Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.
Mirion Technologies (MGPI) SA
Route d’Eyguières
FR-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna, GA 30082
USA
T +1 770 432 2744
F +1 770 432 9179
Mirion Technologies (MGPI H&B) GmbH
Landsberger Strasse 328a
DE-80687 Munich
Germany
T +49 (0) 89515 13-0
F +49 (0) 89515 13 169
Mirion Commercial (Beijing) Co., Ltd.
Shanghai Jiangchang Commercial Branch
Room 801, 78 Jiangchang SanLu
Zhabei District, Shanghai 200436
PR of China
T +86 21 6180 6920
F +86 21 6180 6924
KG 151 RBT/REZ
Detectors
TECHNICAL CHARACTERISTICS
ACCESSORIES
●
Assembly kit with insulated clamps
●
TKS 01: cable protection unit for LOCA-proof
installation of KG 151 REZ
●
TKA 66.11: calibration source Cs-137, remote
operated
●
Connection cables: low noise cable, e.g. Raychem
5021B or Gore GCX 0157
●
Digital signal processing units:
○
TK 250 system: e.g. DPK 251 digital signal
processing channel with c/f-converter NV 103
○
RAMSYS system: LPU-IC or LPDU-IC local signal
processing units
www.mirion.com
144976EN-D
Properties
KG 151 RBT
KG 151 REZ
Tolerance Conditions
Measuring range
1e–7 Gy/h … 1 Gy/h
1 Gy/h … 1e+2 Gy/h
1e–7 Gy/h … 1 Gy/h
1 Gy/h … 1e+2 Gy/h
± 20% acceptable deviation of sensitivity
± 40 % acceptable deviation of sensitivity
Sensitivity
Energy range
2e–6 A/(Gy/h)
80 keV … 3 MeV
2.23e–6 A/(Gy/h)
80 keV … 3 MeV
± 10%
± 40% acceptable deviation of sensitivity (KTA
1501)
Detector housing
Gas fi lling
Embedded source
Steel
Argon, 26 bar
Cs-137, 185 kBq
Stainless steel
Argon, 26 bar
Cs-137, 185 kBq
Operating voltage
Break down voltage
Detector source current
1200 VDC, typ.
2500 VDC
1e-12 A approx.
1200 VDC, typ.
2500 VDC
1e-12 A approx.
± 5%
10 s max.
With embedded source
Temperature range
Short term temperature range
External pressure
Humidity
0 … 150°C (0 ... 302°F)
< 180°C (356°F)
0.7 … 1.3 bar
< 95%
-25 … 150°C (-13 ... 302°F)
< 180°C (356°F)
0.7 … 1.3 bar
< 95%
1 h max.
Length of detector
Total length
Diameter
Sensitive volume
Weight (without connector and cable)
502 mm (19.7 in)
595 mm (23.4 in)
150 mm (5.9 in)
5.7 l
Approx. 12.5 kg (27.5 lb)
502 mm (19.7 in)
602 mm (23.7 in)
150 mm (5.9 in)
5.7 l
Approx. 14 kg (30.8 lb)
Connectors
Fischer size 105
Fischer size 105
Max. cable length is depending on signal pro-
cessing unit (cable capacity)