beautypg.com

Technical data, Example of applications – Mirion Technologies SB 40_150 User Manual

Page 2

background image

Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.

Mirion Technologies (MGPI) SA

Route d’Eyguières

FR-13113 Lamanon

France

T +33 (0) 4 90 59 59 59

F +33 (0) 4 90 59 55 18

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway

Suite 150

Smyrna, GA 30082

USA

T +1 770 432 2744

F +1 770 432 9179

Mirion Technologies (MGPI H&B) GmbH

Landsberger Strasse 328a

DE-80687 Munich

Germany

T +49 (0) 89515 13-0

F +49 (0) 89515 13 169

Mirion Commercial (Beijing) Co., Ltd.

Shanghai Jiangchang Commercial Branch

Room 801, 78 Jiangchang SanLu

Zhabei District, Shanghai 200436

PR of China

T +86 21 6180 6920

F +86 21 6180 6924

SB 40/150

Detectors

TECHNICAL DATA

-

EXAMPLE OF APPLICATIONS

www.mirion.com

145024EN-2

Properties

SB 40

SG 150

with measuring vessel RSD 200,
lead shielding RSA 310 (if applica-
ble), U

discr

= -100 mV

with measuring vessel RSD 275 and
lead shielding RSA 482 RSA 482,
U

discr

= -150 mV

Calibration factor for betas

Measuring range

Background
Gamma sensitivity

Kr-85
Xe-133
Kr-85, 5 cm lead shielding
Xe-133, 5 cm lead shielding
in 5 cm lead shielding
Cs-137

3e4 Bq/m

3

s

-1

±10%

4.4e4 Bq/m

3

s

-1

±20%

5e3 … 5e9 Bq/m

3

1e4 … 1e10 Bq/m

3

0.4 s

-1

8.5 s

-1

/μGy/h, without shielding

4.4e3 Bq/m

3

s

-1

±20%

8.1e3 Bq/m

3

s

-1

±20%

2e3 … 5e8 Bq/m

3

3.37e3 … 1e9 Bq/m

3

2.5 s

-1

0.2 s

-1

/μGy/h, with 5 cm lead

Output signal
Pulse width
Detector voltage
Supply voltage

0 ... -5V at 50 Ohms
0.7 μs approx.
1000 V, <0.15 mA
14.25 ... 25 V, 30 mA approx.

0 ... -5V at 50 Ohms
0.6 μs approx.
1200 V, <0.25 mA
14.25 ... 25 V, 30 mA approx.

Scintillator
Detector housing
Light-tight cover foil
Magnetic shielding

Material; Ø; thickness

KL236; Ø 50 mm; 0.5 mm
Aluminium, Brass
Aluminium, 9 μm
μ-metal

NE102A; Ø 150 mm; 0.5 mm
Aluminium, steel
Aluminium, 12 μm
μ-metal

Connectors
Cables

Fischer S105 A 06093
TKK 05

Fischer S105 A 06093
TKK 05

Temperature range
Ambient pressure
Dimensions

Weight

Ø (Flange) x L without
connector
without connector

0 … +50°C (+32 ... +122°F)
0.7 … 2 bar
65 (90) × 310 mm
(2.5 (3.5) x 12.2 in)
1.5 kg (3.3 lb) approx.

0 … +50°C (+32 ... +122°F)
0.7 … 1.25 bar
150 (232) × 382 mm
(5.9 (9.1) x 15 in)
4.6 kg (10.1 lb) approx.

Beta particulate
monitor AD 24
with SB 40

Noble gas
detector assembly
with SB 40

Detector SB 40 with
Plastic scintillator Ø 50mm

Measuring vessel RDS 200 approx.
1ltr
Lead shielding RSA 310 50mm

Remote check source TKA 61 Sr-90

Remote check source TKA 61 Sr-90