Health physics division, Detection, Detector gamma csi(tl): > 400 cps per µsv/h for – Mirion Technologies PDS 100 G_GN ID User Manual
Page 2: Detector neutron lii(eu) (gn version only), Gamma alarm response time, Identification by nmd algorithm, Up to 4 isotopes mixed, Identification time at 1 µsv/h typical 1 minute, Detection, search and/or identification mode, Source indication alarm and danger alarm

Health Physics
Division
Since norms, specifications and designs are subject to occasional change, please ask for confirmation of the information given in this publication.
Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179
Mirion Technologies (MGPI) SA
Lieu-Dit Calès,
Route d’Eyguières
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
Mirion Technologies (RADOS) Oy
P.O. Box 506
FIN-20101 Turku
Finland
T +358 2 468 4600
F +358 2 468 4601
Mirion Technologies (RADOS) GmbH
Ruhrstrasse 49
DE-22761 Hamburg
Germany
T +49 (0) 40 851 93-0
F +49 (0)40 851 93 256
Mirion Technologies Shanghai Branch
Room 801, 78 Jiangchang
San Lu, Zhabei District
Shanghai 200436
China
T +86 21 6180 6920
F +86 21 6180 6924
PDS-100G-GN/ID
Detection and Identification - Risks Prevention
www.mirion.com
144215EN-C
DETECTION
●
Detector gamma CsI(Tl): > 400 cps per µSv/h for
137
Cs
●
Detector neutron LiI(Eu) (GN version only)
●
Gamma dose rate display 0.01 µSv/h
to 100 µSv/h (1 µrem/h to 10 mrem/h)
●
Gamma count rate display: 0 to 99 999 cps
●
Neutron count rate display: 0.0 to 999 cps
●
Gamma alarm response time:
●
Standard threshold: 0.5 µSv/h step, alarm within 1 s
●
Sensitive threshold: 0.05 µSv/h step, alarm within 3 s
●
Neutron alert response time: mean time to detect
20000 n/s Cf252 at 10 cm: <2 sec.
SEARCH
●
1s integration time with chirp
SPECTROMETRY AND IDENTIFICATION
●
512 / 1024 channels spectra : 30keV to 1.7 MeV
●
Automated alarm triggered mode with auto
confirmation
●
Manual mode with preset time and/or counts and
resume capability
●
Identification by NMD algorithm
○
up to 4 isotopes mixed
○
detectabiliby grade, unknown or ID unsure indication
○
identification time at 1 µSv/h typical 1 minute
○
designed to exceed ANSI N42-48 SPRD standard
FUNCTIONAL FEATURES
●
Detection, search and/or identification mode
●
Source indication alarm and danger alarm
●
Visual, audible and silent alarms (vibration, earphone)
●
Easy-to-read display (OLED technology)
●
Memory of 100/50 512 ch/1024 ch. spectra
and >1000 events
●
IRDA and Bluetooth® technology communication
ELECTRICAL & MECHANICAL CHARACTERISTICS
●
Power supply: 2 x AA batteries
(Lithium, Alkaline or Ni-MH); lifetime typical 100 hours
●
Dimensions (l x w x h): 123 x 74 x 43 mm
(4.84 x 2.91 x 1.69 in); weight: 300 g (10.58 oz)
ENVIRONMENTAL CHARACTERISTICS
●
Temperature range:- 20°C to 50°C (-4°F to 122°F)
ID: -15°C to +45°C (5°F to 115°F)
●
EMI, shock, vibration, drop and water resistant
ACCESSORIES
●
PDSmass software for remote display, spectra and
historic retrieve, parameters settings
●
SMI software for spectra analysis and identification replay
●
Silicon protection
●
Belt Clip
●
Pouch with belt clip
●
External power supply / battery charger
Auto ID in progress
Detection mode screen
Search mode screen
Manual ID in progress
NMD result screen
Spectra view with PDSmass
Auto ID in progress
NORM
40K, 226Ra and daughters, 232Th and daughters
Medical
18F, 51Cr, 67Ga, 75Se, 99mTc, 111In, 123I, 131I,
201Tl
Industrial
22Na, 57Co, 60Co, 133Ba, 137Cs, 152Eu, 192Ir,
241Am
SNM
235U, 238U, 239Pu
In bold, isotopes addressed by ANSI for SPRD