Health physics division, Nuclear characteristics, Detectors – Mirion Technologies HDS-101G_GN User Manual
Page 2: Silicon diode for high gamma dose rate measurement, Lii (eu) scintillator for neutron measurement, Energy range, 30 kev to 3 mev (gamma and x-rays), Measurement and identification range (gamma), Extended gamma range, Three user’s profiles: routine, expert, custom

Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179
Mirion Technologies (MGPI) SA
Lieu-Dit Calès,
Route d’Eyguières
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
Mirion Technologies (RADOS) Oy
P.O. Box 506
FIN-20101 Turku
Finland
T +358 2 468 4600
F +358 2 468 4601
Mirion Technologies (RADOS) GmbH
Ruhrstrasse 49
DE-22761 Hamburg
Germany
T +49 (0) 40 851 93-0
F +49 (0)40 851 93 256
Mirion Technologies Shanghai Branch
Room 801, 78 Jiangchang
San Lu, Zhabei District
Shanghai 200436
China
T +86 21 6180 6920
F +86 21 6180 6924
HDS-101G/GN
Detection and Identification - Risks Prevention
www.mirion.com
145298EN-D
Health Physics
Division
Since norms, specifications and designs are subject to occasional change, please ask for confirmation of the information given in this publication.
NUCLEAR CHARACTERISTICS
●
Detectors:
○
CsI(Tl) scintillator for spectrometry
and low gamma dose rate measurement
○
Silicon Diode for high gamma
dose rate measurement
○
LiI (Eu) scintillator for neutron
measurement
●
Energy range:
○
30 keV to 3 MeV (gamma and X-rays)
○
0.025 eV to 15 MeV (neutrons)
●
Measurement and identification range (gamma):
○
0.01 µSv/h to 100 µSv/h (1 µrem/h to 10 mrem/h)
●
Extended gamma range:
○
Continuous: 0.1 mSv/h to 10 mSv/h, 10 mrem/h to 1
rem/h,
○
Temporarily: up to 1 Sv/h (100 rem/h)
ALGORITHM PROCESSING
●
Continuous spectra acquisition (0.2s time slot) and
stabilization without source
●
Continuous dose rate and count rate comparison to
background
●
Varying Background Suppression algorithm (VBS)
continuously analyses the spectra shape and rejects
alerts due to sudden background changes
●
NORM Medical Discrimination algorithm (NMD)
categorizes and identifies up to 4 isotopes simultaneously
NORM:
40K, 226Ra and daughters, 232Th and daughters
Medical:
18F, 51Cr, 67Ga, 75Se, 99mTc, 111In, 123I,
125I, 131I, 133Xe, 201Tl
Industrial:
22Na, 57Co, 60Co, 133Ba, 134Cs, 137Cs,
152Eu, 192Ir, 241Am, Bremstrahlung
SNM:
233U, 235U, 237Np, 238U, 239Pu
●
Cumulated spectra can be triggered by alert detection or
on request
FUNCTIONAL FEATURES
●
Three operation modes: detection, search and integration
●
Three user’s profiles: routine, expert, custom
●
Alert threshold configurable for dose rate and/or number
of sigma level increase
●
Radiation safety alarm
●
LCD display with backlight
●
Pre-loaded languages
●
Memory to store more than 1000 events/ measurements
and over 100 spectra in different modes
●
Real time data and spectrum transmission
●
Bluetooth®, RS232 and USB interfaces
●
Audio output for earphone
●
Designed to meet/exceed ANSI 42-33, Type I & II and to
IAEA standards
ELECTRICAL & MECHANICAL CHARACTERISTICS
●
Power supply: 6 x AA batteries (Ni-MH rechargeable),
built-in charger
●
Battery life time: 24 hours (Bluetooth disactivated)
●
Dimensions: 280 mm x 78 mm (11 in x 3.1 in)
●
Weight: 1500 g (52.9 oz)
ENVIRONMENTAL CHARACTERISTICS
●
Temperature range: -20°C to 50°C (-4°F to 122°F)
●
Humidity: < 90% at 42°C (108°F)
●
EMI, shock, vibration and drop resistant
●
Water protection (IP54)
●
CE compliant (RF susceptibility 3 V/m)
ACCESSORIES
●
Shoulder strap
●
USB cable
●
Transportation case
●
HDSMASS and SMI
software
●
Power supply/charger/
wall or car cradle
HDSMass software spectra