Page 2: Saturation above 10 sv or 10 sv/h, Linearity, X and gamma energy range: 50 kev to 7 mev, Accuracy hp(10) module alone, Cs : 1 msv/h) (100 mrem/h), Dimensions, Ø: 40 mm (1.6 in)

background image

Health Physics
Division

Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway

Suite 150

Smyrna Georgia 30082

USA

T +1.770.432.2744

F +1.770.432.9179

Mirion Technologies (MGPI) SA

Lieu-dit Calès, route d’Eyguières

F-13113 Lamanon

France

T +33 4 90 59 59 59

F +33 4 90 59 55 18

Mirion Technologies (RADOS) Oy

P.O. Box 506

FIN-20101 Turku

Finland

T +358 2 468 4600

F +358 2 468 4601

Mirion Technologies (RADOS) GmbH

Ruhrstrasse 49

DE-22761 Hamburg

Germany

T +49 (0) 40 851 93-0

F +49 (0)40 851 93 256

Mirion Technologies Shanghai Branch

Room 801, 78 Jiangchang

San Lu, Zhabei District

Shanghai 200436

China

T +86 21 6180 6920

F +86 21 6180 6924

OEM Gamma Sensor

Area and Particulate Monitors - Area Monitor

www.mirion.com

144433EN-C

PHYSICAL CHARACTERISTICS

Measurement range:

Dose: 1 μSv to 10 Sv (0.1 mrem to 1000 rem)

Dose rate: 10 μSv/h to 10 Sv/h (1 mrem/h to 1000 rem/h)

For lower dose rate, please contact Mirion Technologies

Saturation above 10 Sv or 10 Sv/h

Linearity:

<± 15% up to 1 Sv/h (100 rem/h)

<± 25% up to 10 Sv/h (1000 rem/h)

X and Gamma energy range: 50 keV to 7 MeV

Accuracy Hp(10) module alone:

<±5% (

137

Cs : 1 mSv/h) (100 mrem/h)

<±10% (

241

Am: 0.2 mSv/h) (20 mrem/h)

ELECTRICAL CHARACTERISTICS

3.3V 1.2 mA for I2C mode, 0.4 mA for UART mode

MECHANICAL CHARACTERISTICS

Dimensions:

Ø: 40 mm (1.6 in)

h: 4 mm (0.16 in)

Weight: < 5.5 g (0.19 oz)

ENVIRONMENTAL CHARACTERISTICS

Temperature range:

-10°C to +50°C (14°F to 122°F)

Storage: -30°C to +71°C (-22°F to 160°F)

EMC: complies and exceeds standards by a large

margin

FUNCTIONAL CHARACTERISTICS

Alarm report output:

no dose rate alarm or pre-alarm:

-

alarm output signal normally low

dose rate «pre-alarm»:

-

alarm output signal oscillates 0.5 Hz

Dose rate alarm: alarm output signal goes high

programmable threshold for dose rate pre-alarm and

alarm

LED fl ashing is programmable for:

Each 4 or 8 pulses

Each μSv (0.1 mrem)

Each 10 μSv (1 mrem)

Each 100 μSv (10 mrem)

2 Connectors:

one for power supply, I2C or UART serial link, and

alarm

one for digital inputs and outputs

CUSTOMIZATION

Setup can be accessed by authorized user

with OEMMASS software

Health Physics
Division