beautypg.com

Mirion Technologies RAD-60 User Manual

Page 2

background image

www.mirion.com

MGP Instruments Inc

5000 Highlands Parkway

Suite 150

Smyrna Georgia 30082

USA

T +1.770.432.2744

F +1.770.432.9179

MGP Instruments SA

BP 1

F-13113 Lamanon

France

T +33 (0) 4 90 59 59 59

F +33 (0) 4 90 59 55 18

RADOS Technology Oy

P.O. Box 506

FIN-20101 Turku

Finland

T +358 2 4684 600

F +358 2 4684 601

RADOS Technology GmbH

Ruhrstrasse 49

D-22761 Hamburg

Germany

T +49 40 85193 0

F +49 40 85193 256

RAD-60 Personal Alarm Dosimeter

Active Dosimetry

20996027_RAD60_EN_A

TECHNICAL SPECIFICATIONS:

Physical characteristics

Radiation detected:

gamma and X-rays
Detectors:

energy compensated Si-Diode
Measurement range:

- dose: 1 µSv - 9.99 Sv or 0.1 mrem - 999 rem
- dose rate: 5 µSv/h - 3 Sv/h or 0,5 mrem/h - 300 rem/h
Calibration:

better than ±5% (Cs-137, 662 keV at 2 mSv/h), Hp(10)
Energy response:

Hp(10), 60 keV - 3 MeV, better than ±25%, up to 6 MeV, better than ±35%
Dose rate linearity:

better than ±15%, up to 3 Sv/h (300rem/h)
Audible alarms:

- eight separate alarms, sound level typically better than 85 dBA at 30 cm
- integrated dose
- dose rate
- dose overfl ow
- dose rate over fl ow at 3 Sv/h or 300 rem/h
- low battery 1and 2
- defect
Alarm thresholds*:

seven preset values each for integrated dose and dose rate, manually
selectable by push-button
Push-button functions*:

- front panel push-button has the following selectable functions:
- toggle between dose and dose rate display
- switch ON/OFF
- chirp ON/OFF
- reset integrated dose
- change alarm thresholds
- activate battery test

*Note: the active alarm tresholds and confi guration of push button functions
can be changed by using ADR-1 Reader Head in combination with RADOS PC
Software.

Mechanical characteristics

Dimensions: 78 x 67 x 22 mm

Weight: 80 g including battery

Envirionmental Characterstics

Temperature range:

-20 oC to +50 oC operational, humidity up to 90% RH, non condensed

Electrical characteristics

Power supply:

one triple A alkaline cell, life typically 1800 h in background fi eld (dose mode)
Reader communication:

by infrared through bottom part; by using ADR-1 Reader Head in

combination with RADOS PC software