Uncertainty, Health physics division, Functional characteristics selectable units – Mirion Technologies SOR_RF User Manual
Page 2: Alarm levels up to one year battery life under, Normal conditions user selectable display modes, Backlit display, Exhaustive periodic self-testing, Training mode included, 1μgy to 10 gy gamma dose rate measurement range, Relative error of ambient measurement

Mirion Technologies (MGPI) Inc
5000 Highlands Parkway
Suite 150
Smyrna Georgia 30082
USA
T +1.770.432.2744
F +1.770.432.9179
Mirion Technologies (MGPI) SA
BP 1
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
Mirion Technologies (RADOS) Oy
P.O. Box 506
FIN-20101 Turku
Finland
T +358 2 468 4600
F +358 2 468 4601
Mirion Technolgies (RADOS) GmbH
Ruhrstrasse 49
DE-22761 Hamburg
Germany
T +49 (0) 40 851 93-0
F +49 (0)40 851 93 256
Health Physics
Division
Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.
SOR/RF
Active Dosimetry - Dosimeters
www.mirion.com
147349EN-A
FUNCTIONAL CHARACTERISTICS
Selectable units:
●
cGy; cGy/h; mSv; mSv/h; μSv; μS/h;
mrem and mrem/h
4 confi gurable dose and dose rate
●
alarm levels
Up to one year battery life under
●
normal conditions
User selectable display modes
●
Backlit display
●
Exhaustive periodic self-testing,
●
including the detector
Selectable Historical record of
●
measurements and events (3800
steps; 10 s; 1 min ;10 min; 1 h; 24 h)
Data storage in EEPROM
●
(qualifi ed > 10 years without battery)
Low battery (16 h left) pre-alarm and
●
alarm if the battery is removed
Neck lanyard or clip
●
Training mode included**
●
PHYSICAL CHARATERISTICS
Hp(10) dose equivalent measurements:
●
from 50 keV to 6 MeV
Ambient gamma dose measurement range:
●
1μGy to 10 Gy
Gamma dose rate measurement range:
●
from 0.1 μGy/h to 10 Gy/h
Gamma dose rate display:
●
from 1* or 10 μGy/h to 10 Gy/h
Saturation indication (above 10 Gy/h)
●
Relative error of ambient measurement:
●
<± 20 % over the dose measurement range
○
Energy response:
●
<± 20 % in the range 60 keV to 2 MeV
○
<± 50 % in the range 2 MeV to 6 MeV
○
Accredited factory calibration to IEC 17025
●
(COFRAC, procedure #127023)
Cs
○
137
(≈ 25 mSv/h), Am
241
(≈ 35 mSv/h),Co
60
(≈ 9 mSv/h)
conformity scope ±10 % Cs137 with
○
uncertainty
(K=2)
= ± 5 %
MECHANICAL CHARACTERISTICS
Dimensions: 80.4 x 48 x 9 mm (fl at housing)
●
(3.16 x 1.85 x 0.35 in)
Weight: 55 g (1.94 oz)
●
ENVIRONMENTAL CHARACTERISTICS
-20° to + 50°C (-4°F + 122°F) (normal operating range,
●
standard battery 3V LiMnO
2
CR2450)
Resistant to EMP and EMC (radars, GSM, jammers,...)
●
200 V/m pulse modulation, (
MiL-STD-461F RS103)
Resistant to water immersion (IP67: 1 m / 39.3 in),
●
drops, shocks,vibrations, low pressure,
Initial conditions, NBC environmental conditions
○
Complies with the following standards:
●
Meets MIL-STD-810 and MIL-STD-461 requirements
○
Qualifi ed by most of the NATO military laboratories
○
Complies with IEC 1283, ANSI 42-20 and NATO D104
○
NATO number: 6665 14 563 9423
●
NATO supplier n° F8929
Development status: Series production
* LLR : Low level Radiation
**a specifi c confi guration software is requi-
red for the XOM reader
SOR/RF inside the
arm-band pouch
(accessory)
SOR/RF version
to be clipped on a
pocket
SOR/RF worn round
the neck