beautypg.com

Health physics division, According to result editing, Checks per dosimeter, per session – Mirion Technologies IRD-2000 User Manual

Page 2: Calibrations, Result printout, Selective access authorization with password and, Operator traceability general characteristics, Total check length: 2 mins (dose 50 μsv), Total uncertainty (k=2) 7% (dose 50 μsv), Max. check capability

background image

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway

Suite 150

Smyrna Georgia 30082

USA

T +1.770.432.2744

F +1.770.432.9179

Mirion Technologies (MGPI) SA

BP 1

F-13113 Lamanon

France

T +33 (0) 4 90 59 59 59

F +33 (0) 4 90 59 55 18

Mirion Technologies (RADOS) Oy

P.O. Box 506

FIN-20101 Turku

Finland

T +358 2 468 4600

F +358 2 468 4601

Mirion Technolgies (RADOS) GmbH

Ruhrstrasse 49

DE-22761 Hamburg

Germany

T +49 (0) 40 851 93-0

F +49 (0)40 851 93 256

Health Physics
Division

Since norms, specifi cations and designs are subject to occasional change, please ask for confi rmation of the information given in this publication.

IRD 2000

Active Dosimetry - Calibrator

www.mirion.com

144646EN-C

FUNCTIONAL CHARACTERISTICS

Checks response with/without recalibration

Results saved in the database with links to check ses-

sions and irradiator calibration

Date loading and discriminant display on dosimeter

according to result

Editing:

Checks per dosimeter, per session

Calibrations

Result printout

Irradiator calibration assisted by Transfer-Standard

dosimeter with automatic traceable acquisitions and

calculation

Radioactive source decay taken into account

Selective access authorization with password and

operator traceability

GENERAL CHARACTERISTICS

.

137

Cs - 0.37 GBq source

Delivered dose rate: 3 mSv/h approx.

Total check length: 2 mins (dose 50 μSv)

Total uncertainty (K=2) 6% (dose 100 μSv)

Total uncertainty (K=2) 7% (dose 50 μSv)

Capacity: 3 slots for DMC2000 per irradiator module

Max. check capability:

90 dosimeters/hour (1 irradiator module)

150 dosimeters/hour (2 irradiator modules)

5 cm (1.96 in) lead protection

:

External dose rate: < 2.5 μSv/h at 0.3 m (11.81 in)

Lockable shutter

Dimensions (irradiator module):

350 x 443 x 529 mm (13.77 x 17.44 x 20.82 in)

Weight: < 80 kg (176 lb)

HARDWARE SETUP

One or two irradiators module

One control module comprising an LDM2000 reader

and a specifi c extension

DOSICAL software (including database run-time)

1 set of cables

Requires 1 standard PC with Windows 2000, XP or

VISTA plus 1 printer

Information common to dosimeter
batch
Reference dose rates reminder

Operator dialog box

Check results: the three dosimeters are
within tolerances (here, 15 %)

Irradiator calibration results

Detail of irradiator dose rate reference setting using
3 Transfer-Standard dosimeters:

3 measurements are displayed per dosimeter for each reference position.
The software checks the coherence of the reference dosimeter results and
suggests a reference rate per position which is then accepted or rejected
by the operator.
The reference rate history and calibration details are saved.