beautypg.com

Mirion Technologies AMP 200 GM User Manual

Page 2

background image

AMP-200

GM Tube-based Ratemeter and Area Monitor

www.mirion-hp.com

TeChniCAL sPeCiFiCATions:

Radiological Characteristics

• detector: energy compensated GM-tube
• detection range (± 10%): 0.5R/hr to 10,000 R/hr
• energy range sensitivity: 11 cps per R/hr

electrical Characteristics

• power supply: 9 Volts (supplied by a 9 Volt battery located in the meter case)
• optional 9 Volt AC adapter available
• battery life: approximately 50 hours of continuous use

Mechanical Characteristics

• meter dimensions (w x h x d): 6.8 x 11.9 x 1.2 cm (2.7 x 4.7 x 0.5 in)
• probe head length: 16.5 cm (6.5 in)
• diameter: 3.3 cm (1.3 in)
• standard cable length: 7.6 m (25 ft)
• maximum cable length: 106 m (350 ft)

environmental Characteristics

• temperature: -10° to 50°C (15 to 120° F )
• relative humidity: 10 to 95% RH (non-condensing)

5000 highlands Parkway
suite 150
Smyrna Georgia 30082
usA
T +1.770.432.2744
F +1.770.432.9179

BP 1
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18

P.o. Box 506
Fin-20101 Turku
Finland
T +358 2 4684 600
F +358 2 4684 601

Ruhrstrasse 49
D-22761 Hamburg
Germany
T +49 40 85193 0
F +49 40 85193 256

128459F