Mirion Technologies AMP 50 GM User Manual
Page 2

5000 Highlands Parkway
suite 150
smyrna Georgia 30082
usA
T +1.770.432.2744
F +1.770.432.9179
BP 1
F-13113 Lamanon
France
T +33 (0) 4 90 59 59 59
F +33 (0) 4 90 59 55 18
P.O. Box 506
FIN-20101 Turku
Finland
T +358 2 4684 600
F +358 2 4684 601
Ruhrstrasse 49
D-22761 Hamburg
Germany
T +49 40 85193 0
F +49 40 85193 256
DRM-1(D)
Data Radiation Monitoring system
www.mirion-hp.com
TeCHNICAL sPeCIFICATIONs:
Radiological Characteristics
• detector: energy compensated GM-tube
• detection range (± 10%): 10 μR/h to 4 R/h
• energy range sensitivity: 17 cps per mR/hr
electrical Characteristics
• power supply: 9 Volts (supplied by a 9 Volt battery located in the meter case)
• optional 9 Volt AC adapter available
• battery life: approximately 50 hours of continuous use
Mechanical Characteristics
• meter dimensions (w x h x d): 6.8 x 11.9 x 1.2 cm (2.7 x 4.7 x 0.5 in)
• probe head length: 16.5 cm (6.5 in)
• diameter: 3.3 cm (1.3 in)
• standard cable length: 7.6 m (25 ft)
• maximum cable length: 106 m (350 ft)
environmental Characteristics
• temperature: -10° to 50°C (15 to 120° F )
• relative humidity: 10 to 95% RH (non-condensing)
129131C