Gmp-25 mda, Αe: 5,4 mev – Mirion Technologies GMP-25 Pancake Beta Probe User Manual

Page 2

background image

www.mirion.com

www.hp-mirion.com

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway

Suite 150

Smyrna Georgia 30082

USA

T +1.770.432.2744

F +1.770.432.9179

Mirion Technologies (MGPI) SA

Lieu-Dit Calès,

Route d’Eyguières

F-13113 Lamanon

France

T +33 (0) 4 90 59 59 59

F +33 (0) 4 90 59 55 18

Mirion Technologies (RADOS) Oy

P.O. Box 506

FIN-20101 Turku

Finland

T +358 2 468 4600

F +358 2 468 4601

Mirion Technologies (RADOS) GmbH

Ruhrstrasse 49

DE-22761 Hamburg

Germany

T +49 (0) 40 851 93-0

F +49 (0)40 851 93 256

Mirion Technologies Shanghai Branch

Room 801, 78 Jiangchang

San Lu, Zhabei District

Shanghai 200436

China

T +86 21 6180 6920

F +86 21 6180 6924

GMP-25 External Probe

Portable Survey Meters

20996374_GMP25_EN_C

TECHNICAL SPECIFICATIONS:

GMP-25 External Beta Probe

radiation detected: gamma-, beta- and alpha-rays

energy range: gamma > 6 keV, beta

Emax

> 100 keV, and alpha > 2 MeV

detector type: GM tube 7313

pulse rate range: 0 - 50 000 cps

mica window with active area: 15.5 cm2 (6.1 in2); window thickness: 1.5 - 2 mg/cm2

sensitivity: 2.8 cps (170cpm) for 90Sr/90Y uniform source of 0.37 Bq/cm2

equipped with retractable cord, lenght 2000 mm (78.74 in)

temperature range: operation: -25°C to +55°C (-13°F to +131°F); storage: -40°C to +70°C
(-40°F to +158°F)

humidity: 0 - 95% relative humidity, non-condensing

Casing of durable ABS/polycarbonate, lenght 330 mm (13 in), width 70 mm (2.75 in)
(max)

weight: 490 g (1.08 lb)

Th

The GMP-25 probe is equipped with a Binder connector and connected to the mating bulk connector interface

at the bottom of RDS-31 family survey meters. Other version with IKELITE 7 bulk connector and a separate

straight cable (various lengths) available upon request.

GMP-25 MDA

parameter / settings Self Count Rate

BG

Sensitivity

Meas. Time

30 cpm

0.1 μSv/h

300 cpm / μSv/h

tb = 60 s ; ts = 60 s

detectors : GM Tube

source to grid separation 1 cm; Grid open area to surface = 71%

effi

ciency (cps / Bq)

MDA

Am-241,

α

E: 5,4 MeV

> 15%

<

~

0,26 Bq/cm²

C-14 β,
Emean: 50 KeV

> 7 ± 1%

< ~ 0,55 Bq/cm²

Sr-90/Y-90, β,
Emean:0,8 MeV

> 35 ± 4%

< ~ 0,11 Bq/cm²

Cl-36, β,
Emean: 0,27 MeV

> 29 ± 4%

< ~ 0,13 Bq/cm²