beautypg.com

Mirion Technologies RDS 80 User Manual

Page 2

background image

www.mirion.com

Mirion Technologies (MGPI) Inc

5000 Highlands Parkway

Suite 150

Smyrna Georgia 30082

USA

T +1.770.432.2744

F +1.770.432.9179

Mirion Technologies (MGPI) SA

Lieu-Dit Calès,

Route d’Eyguières

F-13113 Lamanon

France

T +33 (0) 4 90 59 59 59

F +33 (0) 4 90 59 55 18

Mirion Technologies (RADOS) Oy

P.O. Box 506

FIN-20101 Turku

Finland

T +358 2 468 4600

F +358 2 468 4601

Mirion Technologies (RADOS) GmbH

Ruhrstrasse 49

DE-22761 Hamburg

Germany

T +49 (0) 40 851 93-0

F +49 (0)40 851 93 256

Mirion Technologies Shanghai Branch

Room 801, 78 Jiangchang

San Lu, Zhabei District

Shanghai 200436

China

T +86 21 6180 6920

F +86 21 6180 6924

RDS-80 Surface Contamination Meter

Portable Survey Meters

TECHNICAL SPECIFICATIONS:

Radiological Characteristics

radiation detected: alpha >2 MeV, beta

Emax

>100 keV, gamma and x-rays from 5 keV to 1.3 MeV

detector: end window GM tube, 1,5-2 mg/cm

2

MICA window active surface area 15,2 cm

2

measurement range: surface contamination 1 to 100 000 cps or 0.01 to 1 000 000

Bq/cm

2

alarm levels: freely adjustable alarm levels for contamination level

cps linearity: ±15% ±1 digit over the range. The Bq/cm

2

display is calculated from cps values using

isotope dependent coeffi

cients

Functional Characteristics

display in either cps or Bq/cm

2

surface activity (Bq/cm

2

) display confi gurable for diff erent isotopes

cps follow-up by audible signal with frequency proportional to activity

visual and audible alarm: user settable for surface activity

histogram capability of up to 480 points with user settable logging interval

backlit display with six large digits

built-in self diagnostics

gamma background reduction for Bq/cm

2

measurements

Built-in infra-red port (IrDA)

Electrical Characteristics

power supply: 2 alkaline batteries IEC LR6/AA size (recommended)

battery life time: at least 2000 hours at normal background with alkaline cells

(more than 1 year under normal operation)

battery alarm: two-step alarm for low battery voltage

electromagnetic compatibility: CE compliant

Mechanical Characteristics

rugged plastic case

GM tube Ø 4.4 cm (1.74”)

dimensions: 78 x 126 x 57 mm

weight: 280 g without batteries, 330 g with batteries

Environmental Characteristics

Temperature range:

-25...+55°C operational (-13... 131°F)

-40...+70°C storage (-40... 158°F)

Accessories

CSW software for parameter setting and histogram readings (can be downloaded free of charge at
www.mirion.com)

• CSW-Confi guration SW full version with calibration key ( item no. 1233-246)
• wrist

strap

• neck

strap

Detector under protective grid

IrDA link

20996034_RDS80_EN_D