TecHnIcaL SpecIFIcaTIonS:
Radiological characteristics
remote detector: GM-tube Zp-1301 and Zp 1201
•
measuring range: 0.05 mR/h to 1000 R/h (0.5 μSV/h to 10 Sv/h)
•
accuracy ± 10 % of reading, within the measuring range
•
energy response: ± 20% at 70 keV to 2 MeV related to 137cs
•
sensitivity: 0.3 cps/mR/h (low range), 17 cps/mR/h (high range)
•
electrical characteristics
power supply: two 1.5 Volt c-type alkaline cells
•
battery life: approximately 70 hours of continuous use (30% less if backlight is on continuously)
•
Mechanical characteristics
meter dimensions (w x l x h): 93 x 148 x 56 cm (3.7 x 5.8 x 2.2 in)
•
probe head length: 226 cm (8.9 in)
•
probe head diameter: 33 cm (1.3 in)
•
weight: 2.55 Kg (7.29 lb)
•
casing: black anodized aluminium
•
environmental characteristics
Temperature:
operation: -10° to 50°c (15 to 120° F )
•
storage: -20° to 60°c (-5 to 140° F )
•
relative humidity: 10 to 95% RH (non-condensing)
•
Telepole with optional Integrated