beautypg.com

Meriam M130 T_C Thermocouple Calibrator User Manual

Page 9

background image

M130 – T/C Calibrator

Instruction Manual Z1017-IR

E.O. 6566 Sept. 2008

www.meriam.com

Page 9 of 21

• ENTER key accepts the menu item to edit and saves the selection. During

the SETUP EDIT state the W UNITS key goes back one increment value,
and the FUNC X key advances one increment value.

• T/C measurements can be recorded to memory either manually or

automatically. See the Record Mode section of this manual for more
information.

• When an over-range/under-range message is displayed, the M130 is

measuring values outside the specified range for a particular T/C type (See
the T/C Measurement Accuracy Table for specified ranges). The red
backlight will illuminate and “----” will be displayed.

• If a thermocouple has not been connected to the unit “T/C OPEN” will be

displayed.



T/C Simulation (Manual)

• When unit is turned on the display defaults to the previous mode with the

last T/C type and Units chosen. To enter the Simulation Manual mode
press the FUNC X key until SIM MANUAL appears, then press the
SETUP key.


T/C Simulation (Auto)

• When unit is turned on the display defaults to the previous mode with the

last T/C type and Units chosen. To enter the Simulation Auto mode press
the FUNC X key until SIM AUTO appears, then press the SETUP key.