M130 – T/C Calibrator
Instruction Manual Z1017-IR
E.O. 6566 Sept. 2008
www.meriam.com
Page 10 of 21
MEAS SETUP: Measure Mode Setup menu structure
T/C TYPE
select ENTER to access submenus
mV
engineering units are mV
B
engineering units are temperature only
E
engineering units are temperature only
J
engineering units are temperature only
K
engineering units are temperature only
N
engineering units are temperature only
R
engineering units are temperature only
S
engineering units are temperature only
T
engineering units are temperature only
OFF TMR
select ENTER to access submenus
30 MIN
shuts off in 30 min without new key stroke
OFF
must use ON/OFF key to shut off
F RECAL
select ENTER to access submenu
ZERO
set zero. See “Zeroing the M130” section
VIEW
view previously recorded values
RECORD
select ENTER to access submenus
MANUAL
press ENTER button to record values
AUTO values
automatically
recorded every 5 seconds
Note:
Main and submenus are continuous scrolling menus. Use W or X to
scroll through the menus.