Meriam M130 T_C Thermocouple Calibrator User Manual
Page 16
M130 – T/C Calibrator
Instruction Manual Z1017-IR
E.O. 6566 Sept. 2008
www.meriam.com
Page 16 of 21
Specifications
Type and Range:
T/C types B, E, J, K, N, R, S, T & milli-volts
Milli-volts range -15mV to +80.00 mV
T/C Measurement & Simulation Accuracy: See Table below
mV Accuracy: 0.005% of reading ±6µV
mV Temp. Effect: ±2 µV /
°C (Factory Ref. Temp. = 23°C)
Cold Junction Effect: ±0.05
°C/°C
Cold Junction Accuracy: ±0.1
°C at 25°C
T/C Resolution:
T/C types E, J, K, N, T: 0.1
°C
T/C types B, R, S: 1
°C
mV: 0.001 mV (1 µV)
Temperature:
Storage = -40
°C to +85°C (-40°F to +185°F)
Operating = -10
°C to +50°C (14°F to +122°F)
Humidity: 10% to 90% Rh
Connection:
Measure: standard mini-T/C connector
Simulate: standard mini-T/C connector
Zero: supplied mini-T/C connector with internal copper short
Maximum Input Voltage: 30VDC
Battery Type:
3 x AA battery, field replaceable.
Battery Operation:
>100
hours continuous use without the backlight for measure
or simulation modes, 1 year shelf life, low battery warning at
approximately 2 hours remaining run time.
Enclosure:
(6.5” Ч 3.2” Ч 1.1”) Polycarbonate, Permanently Static
Dissipative, ESD Protection
Enclosure with Boot:
(6.8” Ч 3.5” Ч 1.3”)