T/C Measurement & Simulation Accuracy Table*
Find T/C type in left column and follow the row across to the temperature range of
interest. Then follow the column up to determine the potential error in degrees C.
T/C
Type
±3.0°
C
±2.0°
C
±1.0°
C
±0.5°
C
±0.4°
C
±0.3°
C
±0.2°
C
±0.1°
C
E
-200 to
-179°C
-179 to
15°C
15 to
1000°C
J
-200 to
-165°C
-165 to
1200°C
K
-200 to
1372°C
N
-200 to
110°C
110 to
1300°C
T
-200 to
-168°C
-168 to
-86°C
-86 to
346°C
346 to
400°C
R
-50 to
27°C
27 to
736°C
736 to
1768°C
S
-50 to
25°C
25 to
1179°C
1179 to
1768°C
B
250 to
295°C
295 to
605°C
605 to
1301°C
1301 to
1820°C
*Notes: 1. Accuracy stated above applies to all T/C measurements and all manual
(non-ramp) T/C simulations.
2. Cold junction effect at 25º C included in above table.