Meriam M202 Precision Absolute Manometer User Manual
Page 31
M2 Series
Instruction Manual 9R68-A
Mar, 2008
www.meriam.com
page 31 of 36
Specifications
Type, Range and Display Resolution:
Absolute Isolated (AI) Type:
17 psia (900 mmHg) – XXX.YY
38 psia (2000 mmHg) – XXX.YY
Accuracy:
M202-AI0017: ±0.02 % F.S. (F.S. = 900 mm Hg)
M202-AI0038: ±0.015 % F.S.* from 0-1000 mm Hg
±0.025% F.S.* from 1000-2000 mm Hg
*F.S. = 2000 mm Hg Absolute
Includes the combined effects of temperature, linearity,
repeatability, hysteresis and resolution.
Warm up time = 5 minutes.
Temperature:
Storage = -40°C to +60°C (-40°F to +140°F)
Operating = -5°C to +50°C (23°F to +122°F)
Media Compatibility:
AI: Absolute pressure sensors for use with gases and liquids
compatible with 316L SS
Pressure Limits:
AI units: 77 PSIA (4000 mm Hg Abs)
Connection:
1/8” female NPT, 316L SS. P1 is the pressure connection.
P2 is not accessible (factory plugged with metal disc)
User must use a wrench on the pressure manifold when
installing user’s 1/8” NPT fitting. Do not tighten the
fitting without using a wrench on the pressure
manifold. Failure to use a wrench on the manifold will
damage the plastic enclosure and void warranty No
torque should be applied to the manifold with respect
to plastic enclosure.