Meriam M202 Precision Absolute Manometer User Manual
Page 17

M2 Series
Instruction Manual 9R68-A
Mar, 2008
www.meriam.com
page 17 of 36
Damp Rate Select
Adjustable exponential type damping is available to steady the 
display when measuring pulsating pressures. The M202 has a range 
of damping rates; 0.1, 0.2, 0.5, 1, 2, 5, 10, or 25 seconds. Damping 
is done by averaging new data from the pressure sensor against 
previously collected data. The microprocessor collects data from 
the sensor every 0.1 seconds. The display updates every 0.5 
seconds, showing the current 0.1 second pressure reading. When set 
at 25 seconds, the display updates every 0.5 seconds with the 
average of the previous 25 seconds readings. Therefore, it takes up 
to 25 seconds from the time pressure is applied until the manometer 
displays the full scale applied pressure. Min/Max display updates 
every 0.1 seconds. 
 
To set the damp rate: 
 
Keystroke
Display
1. Enter Program Mode by 
pressing the PRGM key. 
Top line reads “PROGRAM 
MODE”. Bottom line reads 
“UNITS SELECT”. 
2. Press the
key.
Bottom line reads “DAMP 
RATE SELECT”. 
3. Press the PRGM key.
Top line reads “DAMP RATE 
SELECT”. Bottom line shows 
current value. 
4. Press the
or
keys
until the desire damp rate is 
displayed on the bottom line. 
 
Bottom line shows damp rate 
settings in seconds. 
5. Press the PRGM key.
Top line reads “PROGRAM 
MODE”. Bottom line reads 
“UNITS SELECT”. 
6. Press the
key.
Bottom line reads “EXIT”.
7. Press the PRGM key.
Returns to Measure Mode.