Meriam M202 Precision Absolute Manometer User Manual
Page 13
M2 Series
Instruction Manual 9R68-A
Mar, 2008
www.meriam.com
page 13 of 36
2. Press the PRGM key.
Top line reads “UNITS
SELECT” and bottom line shows
current engineering unit.
3. Press the
or
arrow
key until “USER UNIT
SELECT” is displayed.
Top line reads “UNITS
SELECT”
Bottom line reads “FEET or
METERS”.
4. Press the PRGM key.
Top line reads “ALTITUDE
SELECT”. Bottom line reads
“STANDARD”.
5. Press the
arrow key
once.
Top line reads “ALTITUDE
SELECT”. Bottom line reads
“REF. TO USER”.
6. Press the PRGM key.
Top line reads “VALUE=:
00000000”. Bottom line reads
“CHANGE?: NO”.
7. If the value shown in
step 6 is the correct
altitude, press the PRGM
key to accept and the
key to return to Measure
Mode.
- OR -
If the value shown in step
6 is not the correct
altitude, press the
key
Top line reads “PROGRAM
MODE” and bottom line reads
“UNITS SELECT”.
Top line reads “VALUE=:
00000000”. Bottom line reads
“CHANGE?: YES”.
8. Press the PRGM key.
Example: set altitude to
685 feet above sea level.
Top line reads “USER MODE-
FEET”. Bottom line reads
“00000000”.
9. Press the
or
keys
to set the first digit to “6”.
Using the
key the
provides the character
sequence 0 - 9, ( - )
negative, and (.) decimal
point. The ( - ) sign is used
if your location is below
sea level.
Top line reads “USER MODE-
FEET”. Bottom line reads
“60000000”.