22

Specifications

Specifications

Specifications

Specifications


NIST Traceable Accuracy

Digital: ± 0.025% of FS including all affects of linearity, repeatability, hysteresis and temperature*
Analog: ± 0.035% of FS including all affects of linearity, repeatability, hysteresis and temperature*
The DN0010 sensor accuracy is:

Digital: ± 0.050% of FS

Analog: ± 0.056% of FS

*Temperature Performance: Accuracy statement includes all affects of temperature from -20º to +50º C (-4º to
+122º F)
Update Rate, Single pressure sensor: 14 samples per second
Update Rate, Dual pressure sensors: 7 samples per second

Engineering Units: 32 user selectable units, two user defined units

Pressure Limits

DN sensors: 2x range when pressurized on P1 (H

I

) side only, 150PSI when applied simultaneously to P1 (H

I

) and P2

(LO) sides.

D

I

sensors: 3x range when pressurized on P1 (H

I

) side only, 3x range or 150 PSI (whichever is less) on P2 (LO)

side only 1000 PSI when applied simultaneously to P1 (H

I

) and P2 (LO) sides.

G

I

, C

I

, A

I

sensors: 2x range


Zeroing limits
DN, D

I

, G

I

and C

I

sensors: ±5% FS

A

I

sensors: 100 microns (absolute zero reference)


Media Compatibility
DN sensors:

Non-isolated for clean, dry, non-corrosive, non-

condensing gases only (Brass, 316L SS, Silicon gel)
D

I

sensors:

I

solated for fluids compatible with 316L SS & Viton

®

(standard)

G

I

, C

I

, A

I

sensors:

I

solated for fluids compatible with 316L SS


Digital Communication: Meriam Serial Protocol, DLL or LabVIEW

®


RS-232:
19200 baud (adj.), 8 data bits, 1 stop bit, no parity.

RS-485:

Half duplex, 3-wire (A, B, GND), 19200 baud (adjustable), 1 start bit, 1 stop bit, no parity. Multidrop

addressing with up to 255 devices using Meriam Serial Protocol.

USB: USB 2.0

Analog Outputs

Current: 4–20 mA, 2-wire loop powered or 4-wire systems, minimum loop impedance is 50Ω
Voltage: 0–5 VDC, 4-wire
Switches: 2 each, solid state opto-coupler SPST switches

Rated for 80V DC, 100 mA max.
For use 4-wire 4–20 mA or 4-wire 0–5 VDC only.

Connections
P1 & P2 Pressure Ports:

1

8

” NPT (female)

Flushing Ports: 5/16 – 24 SAE/MS J1926 (316L SS plugs included)
Electrical / Communication

Power: 7 position terminal block (1.3 mm diameter holes suitable for 16 – 25 gauge solid or stranded wire)
RS-232: DB-9 (female) connector or 7 position terminal block
RS-485: DB-9 (female) connector or 7 position terminal block
USB: type B female connector

Analog: 7 position terminal block (two each), 4 position terminal block & USB mini-B