For RS-232 communication use the terminal block for power and the DB-9 connector for communications. Provide 8-
36 VDC
power (20 mA minimum) by connecting to the +supply to the PWR+ terminal. Connect the –supply to the
PWR- terminal adjacent to PWR+. For communications connect the receiving device to the M1500’s DB-9 (female)
serial port (see DB-9 Pin Out Table above) or to the M1500’s RX and TX terminals. Terminal labels follow RS-232
standard TIA/EIA-232.The digital interface draws 20 mA of current or less.
RS
RS
RS
RS-
-
-
-485 Communication
485 Communication
485 Communication
485 Communication
For RS-485 communication use the terminal block for power and communication. Provide 8-36 VDC
power (20 mA
minimum) by connecting to the +supply to the PWR+ terminal. Connect the –supply to the PWR- terminal adjacent to
PWR+. For communications connect the receiving device to the M1500’s A and B terminals. The digital interface
draws 20 mA of current or less.
USB Communication
USB Communication
USB Communication
USB Communication
Connect to the USB type B female connector of the M1500. Use with
high power (500 mA) USB ports or powered USB hubs only.
5
4 3 2 1
9
8 7
6
DB9 Pin Out
RS-232
DCE
1
GND
---
2
RX
Out
3
TX
In
5
GND
---
RS-485 B(-)
RS-485 A(+)
RS-232 RX
RS-232 TX
Vin(8-36V)
GND
Chassis GND