Sponsler, Inc. Model T575N Temperature Compensated Totalizer/Printer




EXAMPLE 5:




From the flowmeter calibration sheet the full-scale flowrate is 898 cycles
per second.
K-Factor (EXAMPLE 3) = 60.759

Maximum Count =

Full-scale flowrate frequency X 60

K-Factor

Maximum Count =

898 x 60

60.759

Maximum Count =

53880

60.759

Maximum Count =

886.78

Maximum Count is not exceeded; no pulses will go into storage.



Model T575N Option ‘IC’ EPROM

POSITION

PRODUCT

LOW TEMP

REF. TEMP

HI TEMP

0

Oxygen

-186.046

-183.11

-167.790

1

Nitrogen

-198.255

-195.930

-181.952

2

Argon

-187.209

-186.046

-167.790

3

CO2

-40.050

-16.388

-6.929

4

N20

-51.087

-34.679

15.384

5

CACO2

-40.051

-25.8333

-6.969

6

MAPP

-17.871

15.384

50.921

7

LPG

-40.050

15.384

49.935

8

Calibrate &

switch test

9

Self test

Temp =

°C