Section 1 technical data, Section 1, Technical data – Hach-Lange NISE sc User Manual
Page 5

5
Section 1
Technical data
Subject to change.
General information
AN-ISE sc
AISE sc
NISE sc
Measurement method
Potentiometric measurement using ion-selective electrodes (ISE)
Ammonium and
potassium, nitrate and
chloride, reference
system
Ammonium and
potassium, reference
system
Nitrate and chloride,
reference system
Measurement range
0
to
1000
mg/L
[NH
4
–N]
0
to
1000
mg/L
[K
+
]
0
to
1000
mg/L
[NO
3
–N]
0
to
1000
mg/L
[Cl
–
]
0
to
1000
mg/L
[NH
4
–N]
0
to
1000
mg/L
[K
+
]
0
to
1000
mg/L
[NO
3
–N]
0
to
1000
mg/L
[Cl
–
]
Precision
5
% of the measured value + 0.2
mg/L
1
Reproducibility
5
% of the measured value + 0.2
mg/L
1
Response time (90
%)
< 3
minutes
(5
to
50
mg/L)
Measurement interval
Continuous
pH range
pH
5 to pH
9
Calibration methods
Sensor code for sensor cartridge, 1 and 2-point value correction or matrix
correction
Power consumption
1
W
Power supply
Via sc controller
Data transfer
Via sc controller
Ambient data
Typical environment
Used in the biological phase of municipal waste water treatment
Storage temperature
Sensor: –20 to 60
°C (–4 to 140
°F)
Sensor cartridge: 5 to 40
°C (41 to 104
°F)
Operating temperature
Air: –20 to 45
°C (–4 to 113
°F)
Sample temperature
+2 to 40
°C (35 to 104
°F)
Maximum flow velocity
< 4
m/s
Maximum sensor immersion
depth/pressure
Can be immersed to a depth of 0.3 to 3.0
m
(1
to
10
ft);
maximum pressure: 0.3
bar
(4.4
psi).
Maximum Compressed air output
during cleaning unit operation
3.1 bar (45
psi)