Operation – Hach-Lange NISE sc User Manual
Page 23

23
Operation
SENSOR MENU
VALUE POINT 2
Enter the values for the 2 point value correction (second point).
Note: The following example shows the input when using a
AN-ISE sc probe for ammonium. When using the AISE sc probe,
the input is the same. When using the NISE sc, it is only possible to
input the nitrate and chloride values.
AN-ISE SC NH4–N
Enter the displayed ammonium value
AN-ISE SC K
Enter the displayed potassium value
LAB NH4–N
Enter the laboratory ammonium value
ENTRY COMPLETE
Confirm the values entered
CORR-RESULT
Display the correction results
FURTHER CORR.
Other matrix correction options
None
No
FURTHER CORR.
is activated
MATRIX 2
A 2-point matrix correction can be performed here
NH4
1
Parameter selection for the MATRIX2 correction.
NO3
1
MEAS CONC 1
Saves the currently measured measurement for the first point
DATE
Displays the date of the current correction of the first point
CONC. LABVALUE 1
Entry and display of reference value for the first point
MEAS CONC 2
Saves the currently measured measurement for the second point
DATE
Displays the date of the current correction of the second point
CONC. LABVALUE 2
Entry and display of the reference value for the second point
HIST. CORR.
Selection of one of the last corrections performed
SENSOR CODE
The sensor code can be activated or entered here
ACTIVATION
Activates the sensor code for the individual channels
NH4 + K
1
Activate the sensor code for ammonium and potassium
NO3 + CL
1
Activate the sensor code for nitrate and chloride
NH4+K NO3+CL
1
Activate the sensor code for ammonium, potassium, nitrate and
chloride
FACTORY CALIBRATION
Activates factory calibration
INPUT
Entry of the sensor code
ENTER CORR.
The laboratory values of the last matrix correction can be changed
ENTER LABVALUE (displayed when MATRIX 1 or
MATRIX 2 is performed)
Enter the laboratory values if MATRIX 1 or MATRIX 2 has been
selected
AMMONIUM
1,2
Entry of ammonium laboratory value
NITRATE
1,3
Enter the laboratory value for nitrate
POTASSIUM
1,2
Enter the laboratory value for potassium
CHLORIDE
1,3
Enter the laboratory value for chloride
ENTRY COMPLETE
Confirm the values entered
CORR-RESULT
Display the correction results
NH4–N
1,2
Displays whether or not the ammonium correction was successful
NO3–N
1,3
Shows whether or not the nitrate correction was successful
K+
1,2
Displays whether or not the potassium correction was successful
CL
1,3
Displays whether or not the chloride correction was successful