LUMIStox 300 Operating Manual
Dr. Bruno Lange GmbH
______________________________________________________________________________
______________________________________________________________________________
Page: 33
3.7 Setting the temperature of the measuring shaft
Before the instrument leaves the factory the temperature control of the measuring shaft
is set to 15
°
C. The temperature control of the measuring shaft can be switched on or
off.
The temperature control can be adjusted. To do this, measure the measuring shaft
temperature with an accurate thermometer. (Introduce about 1 ml water into the
measuring cuvette and place the cuvette in the measuring shaft. Measure the water
temperature after the temperature adjustment.) In the adjustment program, enter the
measured temperature via
calibrates the temperature control so as to reach the required temperature of 15
°
C.
3.8 Setting the screen contrast
The screen contrast can be individually adjusted.
Sequence:
Select the option
Left:
Switch the temperature control on
or off with F2.
Adjust the measuring shaft
temperature with F1.
right:
Set the screen contrast.
options
temp cont srv
temp: 15.1°C on
just off
Kontrast
-
+
meas. :15.0°C
+
-