beautypg.com

10 determining optical density (od) – Hach-Lange LUMISTOX 300 LPV321 User Manual User Manual

Page 24

background image

LUMIStox 300 Operating Manual

Dr. Bruno Lange GmbH

______________________________________________________________________________

______________________________________________________________________________

Page: 24

2.10 Determining optical density (OD)

The LUMIStox 300 measuring instrument incorporates a photometer function for

determining the extinction of a sample as optical density (OD) at a wavelength of 485

nm. The photometer function checks whether preliminary dilution of a strongly coloured

sample is necessary, because the LUMIStox 300 can only carry out colour correction

up to an extinction of 1800 mE. It is also used to assess bacterial growth in the

luminescent bacteria test for chronic toxicity, LUMIS·24·tox (LCK486).

LUMIStox measuring cuvettes (LZP187) are used to carry out the measurements.

Work steps:

1.

Add at least 1 ml water, 2% NaCl solution or any other blank solution to the blank

cuvette.

2.

Add at least 1 ml sample to the sample cuvettes.

Select program mode in the 2nd level

of the main menu.

Perform blank balancing of LUMIStox 300 with blank

cuvette.

Measure sample.

Record or print result.

t e s t O D c h r o n i

r e fe r e n c e c h e c k

p l e a s e w a it

b l a n k - c u v e tt e

b l a n k

s a m p le

m e a s u r e

0 .9 1 E

b l a n k