3.3 Self-test (check)

When the instrument is switched on, and each time an evaluation mode is selected, the

LUMIStox 300 automatically performs a self-test.

In addition the correct functioning of the LUMIStox 300 can be checked with the options

shown below. The result of the check can be documented as a test report on a printer if

the final function to be checked is the printer connection, with (see below). Status

and error messages describe the current operational capability in accordance with ISO

9000 ff.

The self-check is no substitute for regular specialist maintenance of the instrument. You

should contact Dr. Lange Service in Germany (++49/(0)211-5288-0) for all matters

concerning maintenance and a maintenance contract.

mem:

Check memory ( = okay; = defective).

disk:

Check diskette drive ( = okay; = defective).

time:

Check clock.

keyb:

The key that has just been pressed is shown on the display. The check can be

terminated by pressing the key.

LCD:

A symmetrical pattern is displayed. Press any key to terminate this test.

Snd:

An acoustic signal is confirmed by pressing any key.

step:

Check the motor and the resetting of the measuring shaft.

PMT:

Check the photomultiplier.

prn:

Check the printer connection with printout.

EEPr:

Check the EEPROM ( = okay; = defective).

Ser:

Check the serial connection.

input:

Entry of instrument-specific factors by Dr. Lange Service.

options

col. relV test

test

mem disk time

keyb LCD snd

step PMT prn

EEpr ser input

test

test

test