Symptom
Current output
signal
Event
code
Cause
Remedy
Error in measuring
value
Variable
Calibration value incorrect
1. Check FSO in atmospheric air. Hold the transmitter up
in the air with the sensor downwards. Check the value
against DS/EN 25814 (see appendix IIa)
2. Check zero point (see appendix IIb)
3. In the case of large reading error, recalibrate
Display switching
between “Oxygen” and
“End of calibration”
End of calibration phase
Go to “Send command” menu and start measuring, see fi g.
47, p. 36
Flashing bell in display
Depending on
the settings
WO10
The measured value has exceeded the transmitter output setting
Change upper range in “Basic settings”, see fi g. 35, p. 26
WO11
The measured value is below the transmitter output setting
Change lower range in “Basic settings”, see fi g. 35, p. 26
WO15
Life time expired. OXY 1100 should be replaced as soon as
possible
Replace OXY 1100
WO20
The temperature of OXY 1100 is below 0°C (32°F).
The accuracy might be reduced.
Accuracy reduced
WO21
The temperature of OXY 1100 is above 40°C (104°F).
The accuracy might be reduced.
Accuracy reduced
EO50
Calibration not possible.
OXY 1100 is worn or defective
Replace OXY 1100
EO51
Output signal from OXY 1100 is too low.
OXY 1100 is worn or defective
Replace OXY 1100
EO55
The OXY 1100 membrane has been damaged
Replace OXY 1100
EO60
Error in EEPROM
Replace transmitter
EO61
Error in communication between the µ-controller and the A/D
converter or error in the A/D converter
Replace transmitter
EO62
Error in the input circuit
Replace transmitter
EO63
Error in the temperature circuit
Replace transmitter
OXY 4100/3150/4150 transmitter