Fig. 48, Advanced settings – Hach-Lange EVITA OXY User Manual User Manual
Page 39

37
In the “Advanced settings” menu it is possible to 
choose which calibration type to be used and which 
type of start after calibration. Furthermore factors for 
compensation for salinity and relative humidity can 
be keyed in.
Fig. 48
Advanced settings
TR. 0
Enabled
Disabled
Manual start
Automatic start
25
50
125 um
Enabled
Disabled
Mem leakage detect
Enabled
TR. 0
After calibration
Automatic start
TR. 0
Sensor type
50 um
TR. 0
TILTCAL
Enabled
TR. 0
Manual
Automatic
Calibr. type
Manual
TR. 0
00-15
Set transmitter
address 00
10-300 s
Time const.
040. s
TR. 0
High
Low
Normal
High
Low
Hold
Output during
error Low
TR. 0
Output when not
measuring Low
TR. 0
TR. 0
0-100 %
Rel. humidity
060. %
TR. 0
0-40 g/l
Salinity
040. g/l
TR. 0
“Automatic calibration” means calibration in 
atmospheric air and “Manual calibration” means 
calibration in a standard solution where the 
concentration of oxygen is keyed in, see fi g. 41, p. 
32.
“Transmitter address” has to be 0 for current output 
to function in a point-to-point installation using USC 
5000 or 6000. When operating in multidrop system 
using USC 7000, the transmitter must have a 
unique address between 1 and 15.
Advanced settings