4 calibration via foundation fieldbus, Operation—foundation fieldbus – Hach-Lange SI792 P, SI792x P, SI792x P-FF, SI792x P-PA User Manual User Manual
Page 64

Operation—Foundation Fieldbus
62
6.2.4
Calibration via Foundation Fieldbus
The transmitter can be calibrated via Foundation Fieldbus using the
comparison or grab sample method.
pH
1. Make sure the system is configured for pH
(PRIMARY_VALUE_TYPE = pH).
2. Collect a grab sample and set CAL_SAMPLE_PRD_PH to sample.
The pH value of the sample is stored. After writing, the parameter is
automatically reset to NOP (no operation).
3. Read the parameter CAL_SAMPLE_PRD_PH_STORED_VAL. It
contains the stored value.
4. Measure the grab sample and write the lab value in
CAL_PRODUCT_PH. The device is now calibrated.
CAL_SAMPLE_PRD_PH_STORED_VAL is reset to zero.
Note: Calibration values can also be entered directly in the
CAL_SLOPE_PH and CAL_ZERO_PH parameters.
ORP
1. Make sure the system is configured for ORP
(PRIMARY_VALUE_TYPE = ORP).
2. Collect a grab sample and set CAL_SAMPLE_PRD_ORP to
sample. The ORP value of the sample is stored. After writing, the
parameter is automatically reset to NOP (no operation).
3. Read the parameter CAL_SAMPLE_PRD_ORP_STORED_VAL. It
contains the stored value.
4. Measure the grab sample and write the lab value in
CAL_PRODUCT_ORP. The device is now calibrated.
CAL_SAMPLE_PRD_ORP_STORED_VAL is reset to zero.