Operation—4–20 mA/HART
40
5.2.3
Calibration mode configuration
An overview of the calibration setup menu is shown in
for buffer tables.
5.2.4
Alarm settings configuration
An overview of the alarm setup menu is shown in
Table 10
.
Table 9 Calibration mode setup menu—CAL mnu
Select pH buffer set
CA.SOL
–01–BUF
Knick/Mettler-Toledo
–02–BUF
Merck Titrisols, Riedel Fixanals
–03–BUF
Ciba (94)
–04–BUF
NIST technical buffers
–05–BUF
NIST standard buffers
–06–BUF (default) HACH buffers
–07–BUF
WTW technical buffers
–08–BUF
Hamilton Duracal
MAN (manual entry, for buffers not listed in this table)
DAT—Entry of offset and slope of premeasured electrodes
Enter calibration timer interval
CA.tiME
Enter the time interval for calibration (0 to 9999 hours; default: 0 h
to disable timer)
Table 10 Output setup menu for alarm settings
Select Sensocheck
AL.SnSO
CHECK ON
Continuous Sensocheck evaluation of probe
function
CHECK OFF (default)
Enter alarm delay
AL.dLY
0010 sec (default)
Range: 0000–0600 sec
LED in Hold mode
AL.LED
HOLD ON
LED blinks during hold
HOLD OFF
LED off during hold