Replacement parts and accessories – Hach-Lange CLT10 sc User Manual
Page 19

Table 3 Warning list for the sensor (continued)
Warning
Description
Resolution
PH MAINT RECD
pH sensor maintenance is recommended.
The offset is outside of ±45 mV but within ±60 mV.
Clean the sensor and replace the
salt bridge and standard cell
solution and then repeat the
calibration, or replace the sensor.
T MAINT RECD
The temperature offset is outside of ±3 °C but within
±5 °C.
Calibrate the temperature.
Replacement parts and accessories
W A R N I N G
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Parts
Description
Item no.
Cable, gateway to controller, 1 m
6122400
Cable assembly, 4 pin, shielded, flow sensor, 2 m
9160900
Cable assembly, 5 pin, shielded, M12, chlorine sensor, 1 m
9160300
Fitting, inlet port for chlorine flow cell
English:
9159200 (1/4 in.)
Metric:
9196400 (6 mm)
Fitting, speed fit, outlet port for chlorine flow cell or inlet port for pH flow cell
English:
9159300 (3/8 in.)
Metric:
9196500 (10 mm)
Fitting, elbow, outlet port for pH flow cell
English:
9196900 (1/2 in.)
Metric:
9197000 (12 mm)
Float, flow sensor
9159100
Flow cell assembly, chlorine
English: 9198400
Metric: 9198500
Flow cell assembly, pH combination
English: 9180200
Metric: 9180400
Flow cell assembly, pHD
includes sealing hub
English: 9180100
Metric: 9180300
Flow cell body only (pH)
9159700
Flow cell block, chlorine
9158900
Flow control valve
9159000
Flow control valve kit
9198600
Flow sensor
9160800
English 19