Cursor setting – Global Specialties 3600 User Manual
Page 20
Users Manual
Model 3600
www.globalspecialties.com
Page 15
are the same as timing waveform display mentioned in the Timing
Waveform Display section. The functions of window addresses, rolling
steps, find-data words are also identical. The difference is that rolling
up and down in the data listing display equals rolling left and right in
the timing waveform display. The found data is a row of orange data
displayed in the upper part of data listing. The data listing cannot roll
left and right, or zoom.
Cursor Setting
There are two cursors for Model 3600: Cursor1 and Cursor2.
Timing Waveform Cursor
In the timing waveform display interface, press the
【cursor 1】key,
the green cursor1 shows an active state. Turn the knob to enable the
cursor1 to move left and right. Press the
【cursor 2】key, the purple
cursor is then chosen. Turn the knob to enable the cursor2 to move
left and right. The step width of the cursor movement can be set with
number keys 0~9. For example, press
【1】and turn the knob a step.
The cursor moves one point distance on the screen. Press
【9】, then
turn the knob a lattice, the cursor moves nine points distance on the
screen. Recommendation is to first move the cursor with a longer step
to the desired aim point. Then move the cursor with a shorter step to
the aim point.
Cursor Measurement
Model 3600 is able to measure the sampling data value of any point in
the display waveforms and the time interval between any two points in
a waveform.
The second row on the left bottom of the waveform frame is the
parameter value of cursor1. The third row is the parameter value
of cursor2. The six numbers on the left of the parameter are the
decimalization data address that the cursor is indicating. The 8
numbers on the right of the parameter are the Hex data value of the
position the cursor is indicating. This is the sampling data of 32 input
channels. Each number represents 4 channels. The data from left to
right represents the 32 input channels of waveforms from the top to
the bottom in turn. Because the screen only shows the waveforms
of 8 channels, the user can know the logic levels of the 32 channels.
Model 3600 User Manual_B.indd 15
7/2/2010 4:41:45 PM