Brooks, Model 5860 i – Brooks Instrument 5860i User Manual

Page 9

background image

1-3

Brooks

®

Model 5860

i

Section 1 Introduction

Installation and Operation Manual
X-TMF-5860i-MFM-eng
Part Number: 541B110AAG
November, 2008

Ambient Temperature Limits
Operating: 5 to 65°C (40 to 150°F)
Non-Operating: -25 to 100°C (-13 to +212°F)

Working Pressure
4500 psi (31.03 MPa) maximum

Output Signal
0-5 Vdc into 2000 ohms or greater. Maximum ripple 3 mV. Jumper
selectable 4-20 mAdc or 0-20 mAdc. Refer to Figure 2-3 for maximum total
loop resistance.

Temperature Sensitivity
Zero: Less than ±0.075% F.S. per degree C
Span: Less than ±1.0% F.S. shift over 10-50° C range

Pressure Sensitivity
0.03% per PSI up to 200 PSIG

Power Supply Sensitivity
Negligible over the allowable voltage range

Mounting Attitude Sensitivity
±0.5% maximum full scale deviation after re-zeroing

Leak Integrity
1 x 10

-9

Atm. scc/sec Helium

Usable Range
50 to 1

Mechanical Connection
Compatible with most popular mass flow equipment. Refer to Figure 2-1.

Electrical Connection
D-type, 15 pin connector (DA-15P)

*Standard temperature and pressure in accordance with SEMI
(Semiconductor Equipment and Materials International) standard: 0°C and
101.3 kPa (760 Torr). The mass flowmeter can be calibrated to other
reference standard conditions. Specify at time of ordering.