Section 2 Installation

Installation and Operation Manual
X-TMF-5860i-MFM-eng
Part Number: 541B110AAG
November, 2008

2-4 Gas Connections

Standard inlet and outlet connections supplied on the Model 5860

i are 1/

4" compression fittings for flow rates up to 10 slpm, and 3/8" compression
fittings for higher flow rates.

Optional 1/8 compression, VCO

TM

and VCR

TM

connections are available

upon request. Prior to installation, make certain all piping is clean and free
of obstructions. Install the piping in such a manner that permits easy
removal if the instrument is to be removed for cleaning or test bench
troubleshooting.

2-5 Installation (Refer to Figures 2-1 through 2-2)

Recommended installation procedures:

a. The Model 5860

i should be located in a clean dry atmosphere

relatively free from shock and vibration.

b. Leave sufficient room for access to the electrical components.
c. Install in such a manner that permits easy removal if the instrument

requires cleaning.

d. The Model 5860

i Mass Flowmeter can be installed in any position.

However mounting orientations other than the original factory
calibration (see data sheet) will result in a ±0.5% maximum full scale
shift after re-zeroing.

e. When installing flowmeters with full scale flow rates of 10 slpm or

greater, be aware that sharp, abrupt angles in the system piping
directly upstream of the meter may cause a small shift in accuracy.
If possible have at least 10 pipe diameters of straight tubing up-
stream of the Model 5860

i Mass Flowmeter.