Section 4 Maintenance

Installation and Operation Manual

X-TMF-5860i-MFM-eng

Part Number: 541B110AAG

November, 2008

Trouble

Possible Cause

Check/Corrective Action

Output stays at 0 Volts

Clogged Sensor.

Clean sensor. Refer to cleaning procedure. Section 4-2.

regardless of flow.

Output signal stays at +6.8V

Defective PC Board.

Replace PC Board. Refer to Section 4-4.

and there is no flow through
the meter.

Defective Sensor.

Replace Sensor Assembly. Refer to Section 4-4.

Meter grossly out of calibration. Partially clogged sensor.

Clean sensor, refer to the cleaning procedure. Section 4-2.

Flow is higher than indicated.

Meter grossly out of calibration. Partially clogged restrictor.

Replace restrictor. Refer to Section 4-4.

Flow is lower than indicated.

Meter output oscillates.

Anticipate potentiometer out of adjustment. Adjust anticipate potentiometer. Refer to Section 3-4.

Faulty pressure regulator.

Check regulator output.

Defective PC Board.

Replace PC Board. Refer to Section 4-4.

Table 4-1 Bench Troubleshooting

Table 4-2 Sensor Troubleshooting