Ipt122 series – Brooks Instrument IPT122 2 Indicating Pressure Transmitters" User Manual
Page 13
Installation and Operation Manual
X-PR-IPT122-eng
Part Number 541B151AAG
April, 2010
IPT122 Series
2-3
Section 2 Installation
2-8 IPT122 Indicating Pressure Transmitter Mechanical Installation
!
Failure to follow these procedures may adversely affect the product’s
performance and could void the product warranty. Inspect but DO NOT
unwrap any parts until installation. Contact your Brooks representative with
any problems.
1.
Unpack the pressure transmitter
Determine that the desired IPT unit is listed on the packaging. Do not open
if the package does not contain the proper device.
The IPT is double-bagged for cleanroom service and should remain
packaged until installation.
Do not remove the pressure transmitter from the protective bag unless you
are in a clean environment.
A.
Remove pressure transmitter from the box and carry it into the gray
area.
B.
Remove and discard the outer protective bag.
C.
Carry the pressure transmitter (sealed in the inner bag) into the clean
area.
2.
Install the pressure transmitter
A. Prepare the connection fitting in place on the gas line. Any other fitting
components, such as stainless steel gaskets, should be blown clean
with a filtered gas before use.
B. Note: For installation in tight spaces, and on some regulators, it
may be necessary to remove the electronic assembly (back half) of
the IPT. This reduces the depth of the gauge by approximately 7/16".
C. Maintain a flow of at least 1 slpm (0.05 scfm) of inert gas during
installation to minimize tubing and IPT contamination from
environmental moisture and particles. The recommended purge gas
is electronic-grade Nitrogen.
D. Open the inner bag and remove the pressure transmitter. Remove
any fitting protection caps and seat the pressure transmitter on the
mating connections.
E. Tighten the nuts by hand, then 1/8 turn past hand-tight by wrench for
fittings compatible with VCR
®
fittings. DO NOT overtighten.
Note: Refer to specific technical guides furnished by fitting manufacturers for
additional specifications.