Ipt122 series – Brooks Instrument IPT122 2 Indicating Pressure Transmitters" User Manual

Page 12

background image

Installation and Operation Manual

X-PR-IPT122-eng

Part Number 541B151AAG

April, 2010

2-2

Section 2 Installation

IPT122 Series

2-4 Return Shipment

Prior to returning any device to the factory, visit the Brooks web site
(www.BrooksInstrument.com) for a Return Materials Authorization Number
(RMA#), or contact one of the locations provided on p. 2-1.

Prior to returning the device, it must be purged in accordance with the
following:

All devices returned to Brooks require completion of Form RPR003-1,
Brooks Instrument Decontamination Statement, along with a Material
Safety Data Sheet (MSDS) for the fluid(s) used in the instrument. Failure
to provide this information will delay processing by Brooks personnel.
Copies of these forms can be downloaded from the Brooks website
(www.BrooksInstrument.com) or are available from any of the Brooks
Instrument locations provided on p. 2-1.

2-5 Transit Precautions

To safeguard against damage during transit, transport the device to the
installation site in the same container used for transportation from the
factory, if circumstances permit.

2-6 Removal from Storage

Upon removal of the device from storage, a visual inspection should be
conducted to verify its "as-received" condition. If the device has been
subject to storage conditions in excess of those recommended (refer to
Section 2-3 “Recommended Storage Practice" on p. 2-1), it should be
subjected to a pneumatic pressure test in accordance with applicable
vessel codes.

2-7 Gas Connections

Prior to installation, ensure that all piping is clean and free from
obstructions. Install piping in such a manner that permits easy access to
the device if removal becomes necessary.