Power cord-----------------------------------------------------------------1
Spare Fuse-----------------------------------------------------------------1
2. INSTALLATION
2-1. Initial Inspection
This instrument was carefully inspected both mechanically and electrically
before shipment. It should be physically free of damage. To confirm this, the
instrument should be inspected for physical damage in transit. Also, check for
supplied accessories.
2-2. Connecting AC Power
2-2. Connecting AC Power
This instrument requires 115V AC or 230V AC (50-60 Hz) power socket with
protective earth contact (PE-contact). If it is available only a power socket
without PE-contact (so a 2-conductor ac power) then a power socket with PE-
contact must be installed before or the appliance should be provided with a PE-
screw-terminal which is not soluble by hand and not soluble before 2-conductor
power lines.
CAUTION
AC POWER OF THIS INSTRUMENT IS PRESET TO 115 V IN FACTORY.
BEFORE POWERING ON THIS INSTRUMENT, CHECK AND MAKE SURE
THE VOLTAGE OF THE POWER SOURCE IS SAME WITH THE MARKING OF
UNIT.
IN CASE OF 230 V AC, VOLTAGE SELECTOR (on rear panel) SHOULD BE
SELECTED TO 230 V POSITION AND THE FUSE SHOULD BE CHANGED
ACCORDING TO THE FUSE RATINGS IN THE SPECIFICATIONS.
2-3. Cooling and Ventilation
No special cooling and ventilation is required. However, the instrument should
be operated where the ambient temperature is maintained.
2-4. Position
This instrument is built as a bench-type instrument with rubber feet and tilt stand
in place. Stand-up angle can be adjusted by rotating angle of carrying handle.
2-5. Warming-Up