5. Specifications

INSULATION
Test Voltage

250Vdc

500Vdc

1000Vdc

+10%-0%

+10%-0%

+10%-0%

Mirror scale

100MW

200MW

400MW

Mid Scale

1MW

2MW

4MW

Scale Multiplier

x1/2

x1

x2

Accuracy

±3%

Output Short-Circuit Current

1.3mA

Regulated Output Voltage (up to 1mA current)

263.5V / 525V / 1052V

CONTINUITY
Low W

0 - 3W

0 - 500

W

Test Leads / Fuse Zero W Adjustment by knob
Output Short-Circuit Current

205mA

Accuracy

±1.5% of scale length

AC Voltage
Range

0 - 600Vac

Accuracy

±3% of scale length

Voltage Warning
Warning Light Circuit Live lit from 90Vdc / 70Vac
Buzzer Beep from 24Vac / dc

Battery Check
Battery Check indicate good batteries from 8Vdc to 13Vdc
during a Load test of 205mA.

Battery OK
Battery OK Led lit from 8Vdc and is operative while testing.