beautypg.com

Specifications (cont.), Period, Supplemental – B&K Precision 5492BGPIB - Datasheet User Manual

Page 5: Frequency

background image

Period

Specifications (cont.)

5

w w w.bk prec ision.c om

True RMS Bench Multimeter

Model 5492B

v041014

Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C

ACV Range

Period Ran ge

Res olution

Fu ll Scale

Re ading

Accuracy

(2 )

I nput Se nsitiv ity

(Sin e Wave )

100 mV

to

750 V

1 ~ 10 µs

(1)

0.01 ns

9.99999

0.005

100 mVrms

10 µs ~ 10 ms

0.1 ns

9.99999

0.005

40 mVrms

10 ms ~ 100 ms

0.1 µs

99.9999

0.01

40 mVrms

100 ms ~ 200 ms

1 µs

199.999

0.05

200 mVrms

(1) If testing frequency is greater than 1MHz, it will be displayed but no specified accuracy is guaranteed.
(2) Specified accuracy at input >5 % of range.

Supplemental

Environmental

Operating Environ men t

0 °C to 40 °C, 32 ºF to 104 ºF

Storage En vironment

-40 °C to 70 °C -40 ºF to 158 ºF

Hu midity

90 %RH

General

Remote Interfaces

RS232, USB (virtual COM) GPIB (model 5492BGPIB only)

Measu reme nt Category

CAT II 300V, CAT I 1000V, pollution degree 2

Math Fu nctions

Rel, Max/Min/Average/StdDev (for stored readings), dB, dBm, Limit Test, % and mX+b

AC Input

110/220 V ± 10%, 50/60 Hz ± 5%

Powe r Cons umption

20 VA

Safety & EMC

EN 61010-1:2001, EN61326-1:2006 and

EN61326-2-1:2006

Dimen sions (W×H×D )

8.85” x 3.93 “ x 13.97”

(225 mm×100 mm×355 mm)

Ne t Weig ht

5.51 lbs (2.5 kg)

Three-Year Warranty

I ncluded Acce ssorie s

Test leads, AC power cord, spare fuse, user manual, USB cable, calibration certificate

and test report

Frequency

Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C

A CV

Range

Fre qu ency Range

R esolu tion

Fu ll Scale

Re ading

Accuracy

(2 )

Input Sensi ti vity

(Sine Wave )

100 mV

to

750 V

5 ~ 10 Hz

10 µHz

9.99999

0.05

200 mVrms

10 ~ 100Hz

100 µHz

99.9999

0.01

40 mVrms

100 ~ 100 kHz

1 mHz

999.999

0.005

40 mVrms

100 k ~ 1 MHz

(1)

1 Hz

999.999

0.005

100 mVrms

(1) If testing frequency is greater than 1MHz, it will be displayed but no specified accuracy is guaranteed.
(2) Specified accuracy at input >5 % of range.

Gate Time at 1 Sec.
Input protection: 750 Vrms or 500 Vdc

Gate Time at 1 Sec.
Input protection: 750 Vrms or 500 Vdc