5
Models
2511
2512
2515
2516
Performance Characteristics
Bandwidth
60 MHz
100 MHz
60 MHz
100 MHz
Real Time Sampling Rate
1 GSa/s (half-channel interleaved)
(1)
, 500 MSa/s (per channel)
Channels
2 non-isolated
2 isolated
Rise Time
< 5.8 ns
< 3.5 ns
< 5.8 ns
< 3.5 ns
Ch-to-Ch Isolation (both
channels at same V/div setting)
> 100:1 at 50 MHz
Memory Depth
40 kpts (half-channel interleaved)
(1)(2)
, 20 kpts (per channel)
Deep Memory
(3)
2 Mpts (half-channel interleaved)
(1)
, 1 Mpts (per channel)
Vertical Resolution
8 bits
Vertical Sensitivity
2 mV/div – 100 V/div (1-2-5 order)
5 mV/div – 100 V/div (1-2-5 order)
DC Gain Accuracy
5 mV/div-100 V/div: ≤ ± 3 %
2 mV/div: ≤ ± 4 %
5 mV/div-100 V/div: ≤ ± 3 %
Max. BNC Input Voltage
CATII 300 Vrms from BNC signal to BNC shell
Max. Input Voltage for Probe
PR150SA
(4)
: 1x/10x CAT II 300 V
PR250SA
(5)(6)
: 10x CAT III 600 V, CATII 1000 V
Channel Voltage Offset Range
2 mV - 200 mV : ±1.6 V
206 mV - 10 V : ±40 V
10.2 V - 100 V : ±400 V
5 mV - 200 mV : ±1.6 V
206 mV - 10 V : ±40 V
10.2 V - 100 V : ±400 V
Bandwidth Limit
20 MHz (-3 dB)
Horizontal Scan Range
5.0 nS/div - 50 S/div
2.5 nS/div - 50 S/div
5.0 nS/div - 50 S/div
2.5 nS/div - 50 S/div
Timebase Accuracy
± 50 ppm measured over 1 ms interval
Input Coupling
AC, DC, GND
Input Impedance
1 MΩ +/- 2 % || 18 pF ± 3 pF
1 MΩ +/- 2 % || 16 pF ± 3 pF
Probe Attenuation Selectable Factors
1X, 5X, 10X, 50X, 100X, 500X, 1000X
Vertical and Horizontal Zoom
Vertically or horizontally expand or compress a live or stopped waveform
I/O Interface
USB
USB host port support USB flash drives (FAT format)
mini-USB device port for PC connectivity and probe compensation
Acquisition Modes
Sampling
Display sample data only
Peak Detect
Capture the maximum and minimum values of a signal
Average
Waveform averaged, selectable from 4, 16, 32, 64, 128, 256
Trigger System
Trigger Types
Edge, Pulse Width, Video*, Slope, Alternative
*Support signal formats: PAL/SECAM, NTSC
Trigger condition: odd field, even field, all lines, or line number
Trigger Modes
Auto, Normal, Single
Trigger Coupling
AC, DC, LF reject, HF reject
Trigger Source
CH1, CH2
Trigger Level Range
± 6 divisions from center of display
Trigger Displacement
Pre-trigger: Memory depth / 2* sampling
Delay Trigger: 268.04 div
Pulse Width Trigger
Positive slope (>, <, =), Negative slope (>, <, =), Time: 20 ns - 10 s
Slope Trigger
Positive slope (>, <, =), Negative slope (>, <, =), Time: 20 ns - 10 s
Alternate Trigger
CH1 trigger type: Edge, Pulse, Video, Slope
CH2 trigger type: Edge, Pulse, Video, Slope
Digital Storage Oscilloscope Specifications