2 utility key – B&K Precision 4013B - Manual User Manual
Page 13
![background image](/manuals/320592/13/background.png)
13
The offset voltage has three ranges as follows:
Output Voltage
Range
Constraints of Amplitude +
Offset
1.01 volt to 10.00 volts
(Vp-p)/2 + |offset| <= 5 volts
0.101 volt to 1 volt
(Vp-p)/2 + |offset| <= 0.5 volts
0.010 volt to 0.100 volt
(Vp-p)/2 + |offset| <= 0.05 volts
F4:SYM – (Symmetry) When the Square or Triangle waveforms are selected, the
SYMMETRY(Duty cycle) parameter is available. Change the symmetry(Triangle) or duty
cycle(square) by using the cursor keys, rotary dial or numerical keys. If a certain setting
cannot be produced, the waveform generator will display a warning message.
Square Menu
3.6.2 UTILITY Key
Utility Menu
F1: RECALL – Recalls a previously stored front-panel setup from the selected buffer. Change
the buffer number by using the rotary input knob. Valid storage buffer numbers are from 1 to
9.
Buffer 0 is the factory default setup.
F2:STORE – Stores the current front-panel setup to the specified storage buffer. Change the
buffer number by using the data keys or the rotary input knob. Valid storage buffer numbers
range from 1 to 9.
F3: OUT-ON – Selects the OUTPUT state from a power-up. Select ON to enable or OFF to
disable the output from a power-up.