Time interval analyzer & universal counters, Ta320, Recorders – Atec Yokogawa-TA320 User Manual
Page 5: Index
![background image](/manuals/320066/5/background.png)
13
TIME INTER
V
AL ANAL
YZER & UNIVERSAL COUNTERS
RECORDERS
WAVEFORM
MEASURING
INSTRUMENTS
DIGITAL
MEASURING
INSTRUMENTS
INDEX
TIME INTERVAL ANALYZER & UNIVERSAL COUNTERS
TA320
Application Software Specifications
●
Optional Disk Analysis Software for the TA320
Readable data:
Binary type data measured with the TA320(WVF), Binary data
saved using this software
Data acquisition:
Via floppy disk or communications through GP-IB interface
Display of analysis results: Histogram display (graph display), vertical scale: log or lin-
ear, Delta average table, Jitter value (standard deviation) table,
Judgment line display, and Independent/overlapping display for
Pit/Land measured values
Analysis range:
Start cycle is limited to 1 to 3T.
End cycle ranges from 1 to 16T.
Save/load functions: Analysis data and the software settings can be saved or loaded.
Saving or loading of average values only is also available.
Database function:
Data for each Pit/Land and the title to be given to the data can be
controlled as a bundle.
Printing function:
Histograms, graphs, and tables can be printed out.
The printing direction can be specified.
Judgment function:
This function sets the median and allowable range for a delta av-
erage and the criterion for standard deviation.
With the judgment function, the judgment line is displayed on
the graph and the data exceeding the criterion (NG) are marked
in a different color.
Comments:
Comments can be attached to the analysis result.
Personal computer:
Those running Windows 3.1, Windows 95, or Windows NT
workstation
GP-IB board:
Those made by NATIONAL INSTRUMENTS.
■
BLOCK DIAGRAM OF THE TA320