Video Measurement • www.tektronix.com/video

9

Automated Video Measurement Set

VM5000

measurement and high frequency timing

measurements. Utilizing RS-232 control,

the VM5000 automatically selects either

low frequency or wideband mode as

required by the parameter being measured.

Incorporating a full 1 GHz of bandwidth,

with optimized return loss in wideband

mode, the MIU delivers un-matched speed,

accuracy and convenience in testing PC

graphics signals.

Standard GPIB Remote Control
and LAN Connectivity

A fast and reliable GPIB Port compliant to

IEEE 488.2 is standard on the instrument

with selectable controller or talk/listener

modes. A fully documented oscilloscope

GPIB remote command set and simplified

video command set enable all of the

instrument capabilities accessible via the

user interface to be automated via GPIB

remote control.

Network connectivity is provided with

a LAN port supporting 10Base-T and

100Base-T. This enables video test reports

or data stored on the hard drive to be

accessed via the network. Network remote

control can be accomplished via the

Tektronix AD007 GPIB to LAN adapter

accessory, available separately. TekVISA

is functional for LAN remote control of the

oscilloscope commands.

Complete Oscilloscope
Functionality

Recognizing the need for flexibility,

Tektronix has integrated complete digital

phosphor oscilloscope functionality into

the VM5000. Flexible and dedicated

video triggering for NTSC, PAL, SECAM

and analog HDTV is provided in the

oscilloscope. Optional application software

packages complement the standard

oscilloscope functionality to further extend

the return and value of a VM5000 platform

investment. Oscilloscope functionality

and specifications are detailed in the

TDS5104B or appropriate application

software data sheet(s).

Options SD and HD Video Measurements

Characteristic

Description

Color Bar Measurement Accuracy

±2 mV ±1.5% of reading

Measurement of all 8 bar levels, displayed in absolute

(mV) values. YPbPr or RGB formats

Noise Measurement

Range and Accuracy

Unweighted and Weighted

–20 to –60 dB ±1 dB

Specification requires 30 MHz bandwidth filter

–60 to –70 dB ±2 dB

setting. Trigger on CH 4 required for measurements

below –60 dB. Noise measurement bandwidth select-

able, 200 KHz to 250 MHz. Unified Weighting Filter

Noise Floor

<–76 dB, 30 MHz noise bandwidth

Typically <–80 dB, 30 MHz noise bandwidth

Multiburst and Frequency Response Measurements

Accuracy

Flag Amplitude

±2 mV ±1.5% of reading

1 MHz to 10 MHz Packets

±0.5 dB

Measurement relative to reference flag amplitude

10 to 30 MHz Packets

±0.75 dB

Frequency Readout

±0.1 MHz

Multiburst packet frequency is measured

and displayed

Non-linearity Measurement

Accuracy

Non-linearity

±1.5%

Non-linearity measurement using a ramp test signal.

Incremental linearity reported over five equally
spaced intervals, as well as an overall linearity

figure for each channel

Transient Response Measurement

Accuracy

Rise Time and Fall Time

±5 ns

K2T

±1%

Overshoot and Undershoot

±1.5% of reading

Rise and Fall Settle Times

±5 ns

*

1

For VM5000 Instrument Characteristics, please refer to the TDS5104B datasheet.

Characteristics

*1

Video Measurement
Specifications

Oscilloscope Measurement Menu.

chart continued on next page