Safety standards and certification – Atec Tektronix-VM100-VM101 User Manual
Page 6
![background image](/manuals/320049/6/background.png)
Appendix A: Specifications
A–6
VM 100 & VM101 User Manual
Table A–6: Environmental characteristics
(cont.)
Category
Description
Package Drop
REQ:
Meets Tektronix 062-2858-00, Rev. B
Random Vibration
(Operating and Non-Operating)
REQ:
Meets Tektronix Class III (062-2858-00,
Rev. B)
Shock (Operating)
REQ:
Meets Tektronix Class III (062-2858-00,
Rev. B)
RI: Three drops on each face of 50 g,
11 ms duration, half-sine pulse
shape for a total of 18 drops
Storage Temperature
REQ:
–40
°
C to + 75
°
C (–40
°
F to +167
°
F)
Loose Load Vibration
(Repetitive Shock)
REQ:
Meets Tektronix 062-2858-00, Rev. B
(page 51, section 17.7.4)
Vehicle Vibration
REQ:
Meets Tektronix 062-2858-00, Rev. B
Safety Standards and Certification
Table A–7: Certifications and compliances
Category
Description
Altitude (Maximum Operating)
See page A–5 for additional
information
Operating: to 6,500 f (2,000 m)
Certifications
US Nationally Recog-
nized Testing Laboratory
Listing
ANSI/ISA S82
Safety Standard for Electrical and Electronic Test,
Measuring, Controlling, and Related Equipment
Canadian
Certification
CAN/CSA C22.2 No.1010.1
Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use