beautypg.com

Atec Tektronix-VM100-VM101 User Manual

Page 4

background image

Appendix A: Specifications

A–4

VM 100 & VM101 User Manual

Table A–3: Measurement accuracy (cont.)

Category

1

Description

Hum

2

REQ:

±

1.0% for hum

4.0%

±

2.0% for hum >4.0% and <15.0%

RI: Typically

±

0.5% for hum

4.0%

Signal-to-Noise Ratio
(Weighted or Flat)

REQ:

±

1 dB, 28 dB to <45 dB

±

2 dB, 45 dB to <60 dB

±

3 dB, 60 dB to 67 dB (WGT)

RI:

VM100: Noise is weighted with EIA
RS-250-B 5 MHz weighted noise
filter. (Tektronix part number
015-0215-00 equivalent.)

VM101: Noise is weighted with
CCIR 567 weighted noise filter

Group Delay

REQ:

<

20 ns

RI: 50 kHz to 4.8 MHz, within the

bandwidth of the signal under test

Sync Amplitude

2

REQ:

<

1.0%

Burst Amplitude

REQ:

<

1.0%

Bar Amplitude

REQ:

<

1.0%

Luminance Nonlinearity

REQ:

<

1.0%

1

All tests are performed with AVG enabled. Processing time for each
measurement is a function of the signal-to-noise ratio.

2

The hum and sync amplitude measurements are inaccurate in the presence
of Sound In Sync (SIS) signals.