Atec Tektronix-DDS200 User Manual
Appendix a: specifications
DDS200 Digital Demodulation System User Manual
A–1
Appendix A: Specifications
This section contains the specifications for the DDS200 Digital Demodulation
System. All specifications are guaranteed unless labeled “typical.” Typical
specifications are provided for your convenience and are not guaranteed.
Specifications labeled with the
n symbol are checked in the Performance Test
(PT) procedure in the service manual.
NOTE. Performance requirements are valid provided that the instruments are
operating within environmental parameters and have warmed up for at least one
hour.
Table A–1: Electrical Specifications
Characteristics
Requirements
Supplemental information
Overvoltage category
CAT II
RF Input
Connector: front-panel female type BNC
n
Frequency range
47 to 862 MHz
Level
100
mV to 700 mV
Accuracy, typical
" 3 dB
Impedance, typical
75
W
n
Return loss
y 12 dB
Channel bandwidth, typical
8 MHz
IF Input
Level, typical
10 to 100 mV
(80 to 100 dBuV)
Impedance, typical
50
W
n
Return loss (33 to 40 MHz)
y 20 dB
IF Output
Level, typical
45 mV (93 dBuV)
Impedance, typical
50
W
Frequency, typical
36 MHz
Demodulator
n
Modulation modes
1
4, 16, 32, 64, 128,
256 QAM
Pulse response roll-off factors
0.15, 0.20, 0.25,
0.30
1
Operation at 4, 128, and 256 QAM is not specified.