Measurement accuracy - power
Watts Accuracy
(V
RMS
acc. x A
RMS
x PF) ± (A
RMS
acc. x V
RMS
x PF) ± (V
RMS
x A
RMS
x cos θ -cos { θ± (Vh1
ph.err.
± Ah1
ph.err.
)})
VA Accuracy
(V
RMS
acc. x A
RMS
) ± (A
RMS
acc. x V
RMS
)
VAR Accuracy
PF Accuracy
Cos θ -cos [ θ± ( Vh1
ph.err
± Ah1
ph.err
)] ± 0.001
Measurement accuracy - Harmonic
magnitude and phase
1
Voltage Harmonics (45 Hz -
850 Hz) (Typical for 10 Hz to
45 Hz and 850 Hz to 1 MHz)
± 0.08% of Reading ± 0.08% of Range ± (0.02*F)% of Reading ± 0.02 V
Voltage Harmonics Phase
(45 Hz - 850 Hz) (Typical for
10 Hz to 45 Hz and 850 Hz to
1 MHz)
0.025 ± [ 0.005 * V
range
/ V
reading
)] ± (0.1 / V
range
) ± ( 0.002 *F )
Current Harmonics (45 Hz -
850 Hz) (Typical for 10 Hz to
45 Hz and 850 Hz to 1 MHz)
± 0.08% of Reading ± 0.08% of Range ± (0.02*F)% of Reading ± (20 µV /Z
ext
)
Current Harmonics Phase
(45 Hz - 850 Hz) (Typical for
10 Hz to 45 Hz and 850 Hz to
1 MHz)
0.025 ± [0.005 * A
range
/ A
reading
)] ± (0.001 / A
range *
Z
ext
)) ± (0.002 *F)
Physical characteristics
Dimensions
Height
13.2 cm.
5.2 in.
Width
42 cm.
16.5 in.
Depth
31 cm.
12.5 in.
Weight
Net (without lead set)
8.8 kg
19.5 lb.
Temperature
Storage
-20
○
C to +60
○
C
Operating
0
○
C to +40
○
C
1
F is the frequency measured in kHz. In the case of harmonics, F is the harmonic frequency.
Z
ext
is the shunt impedance (0.6 Ω for 1 A shunt, 9.375 mΩ for 30 A shunt). It is assumed that the waveform measured is a sine wave.
Measurement conditions during calibration: Instrument default settings unless otherwise stated, Sine waves applied to V and I inputs, 30 minute warm-up, Temperature 23 °C ±5 °C.
PA4000 Power Analyzer